model | 7630 |
input |
voltage | 115/230vac±15% |
frequency | 50/60hz±5% |
fuse | 2a slow-blo 250v ac |
line condition |
power switch | reverse polarity switch for normal condition (on / off / auto setting) |
neutral switch | neutral switch on/off selection for single fault condition |
ground switch | ground switch on/off selection for class i single fault condition |
probe setting |
surface to surface (ph-pl) |
surface to line (ph-l) |
ground to line(g-l) |
ground to neutral (g-n) |
auto function (g-n & g-l) |
setting |
touch current high limit (rms) | range: | 0.0ua~ 999.9ua/ 1000ua~5999ua/ 6.00ma~20.00ma |
resolution: | 0.1ua/1ua/0.01ma |
touch current low limit (rms) | range: | 0.0-999.9 ua |
resolution: | 0.1ua |
touch current high limit (peak) | range: | 0.0ua - 999.9ua/ 1000ua - 9999ua/ 10.0ma - 30.0ma |
resolution: | 0.1ua/1ua/0.1ma |
touch current low limit (peak) | range: | 0.0-999.9 ua |
resolution: | 0.1ua |
display |
touch current display (rms) | range 1 | 0.0ua~ 999.9ua, frequency dc - 1mhz |
resolution | 0.1ua |
accuracy | dc : ±2% of reading + 3counts 15hz < f <100khz : ±2% of reading + 3counts 100khz < f < 1mhz : ±5% of reading (10.0ua - 999.9ua) |
range 2 | 1000ua~8500ua, frequency dc - 1mhz |
resolution | 1ua |
accuracy | dc : ±2% of reading + 3counts 15hz < f <100khz : ±2% of reading + 3counts 100khz < f <1mhz : ±5% of reading (10ua - 8500ua) |
range 3 | 8.00ma~20.00ma, frequency dc - 1mhz |
resolution | 0.01ma |
accuracy | dc : ±2% of reading + 3counts 15hz < f < 1mhz : ±5% of reading (0.01ma -20.00ma ) |
touch current display (peak) | range 1 | 0.0ua~ 999.9ua, frequency dc - 1mhz |
resolution | 0.1ua |
accuracy | dc : ±2% of reading +2ua 15hz < f <1mhz : ±10% of reading +2ua |
range 2 | 1000ua ~ 8500ua, frequency dc - 1mhz |
resolution | 1ua |
accuracy | dc : ±2% of reading +2counts 15hz < f <1mhz : ±10% of reading +2counts |
range 3 | 8.0ma ~ 30.0ma, frequency dc - 100khz |
resolution | 0.01ma |
accuracy | dc : ±2% of reading +2counts 15hz < f < 100khz : ±10% of reading +2counts |
measuring device modlue |
md1 | iec60990 fig4 u2, iec 60950-1, iec60335-1, iec60598-1, iec60065, iec61010 |
iec60990 fig4 u1, iec60065 |
md2 | iec60990 fig5 u3, iec60598-1 |
iec60990 fig5 u1 |
md3 | iec 60601-1 |
md4 | ul544np, ul484 , ul923, ul471, ul867, ul697 |
md5 | ul544p |
md6 | ul1563 |
md7 | iec60950, iec61010-1 figa.2 (2k ohm) for run test md circuit. |
external md | basic measuring element 1k ohm |
md voltage limit | 70vdc |
leakage current offset | range: | 0 - 999.9ua |
resolution: | 0.1ua |
dut power |
ac voltage | 0.0-277.0v |
ac current | 40arms max continuous |
over current protection | 50arms, response time < 3 sec / 250apeak response time <10u sec |
ac voltage high/low limit | range: | 0.0-277.0v |
resolution: | 0.1v/step |
ac voltage display | range: | 0.0 - 277.0 v |
resolution: | 0.1 volt / step |
accuracy: | ±(1.5% of reading + 2 counts) , 30.0 - 277.0vac |
delay time setting | range: | 0.2 !v 999.9 seconds |
resolution: | 0.1 second |
timer display | range: | 0.0 !v 999.9 seconds |
resolution: | 0.1 second |
accuracy: | ± (0.1% of reading + 0.05 seconds) |
general |
plc remote control | input - test, reset, interlock & recall file 1 through10 |
output - pass, fail, test-in-process, start-out, reset-out |
scanner control | it is applied to control matrix scanner (model 7006) |
memory | 50 memories , 30steps per each memory . max. result display 900 data (30 memories x 30 steps) |
auto reverse function | auto reverse on / off parameter setting selection automatic reverse polarity switch for normal condition in one step setting menu. only display maximum leakage current value. |
scope output interface | at rear panel bnc type to connet scope for some iec standards test requirement and application. |
display | 320 x 240 graphic lcd |
lcd contrast setting | range: 1-9; 1 is lightest character, 9 is darkest character. |
alarm volume setting | range: 0-9; 0=off, 1 is softest volume, 9 is loudest volume. |
security | lockout capability to avoid unauthorized access to test set-up programs. |
calibration | software and adjustments are made through front panel, automatic calibration alert function to signal operator when calibration is due. |
results display | all test result information will be displayed on the screen. |
interface | rs232 standard |
safety | product with tuv/gs certificate |
dimension | 430 (w) x 133(h) x 300 (d) mm |
weight | 12 kg |
standard accessories |
power cord | x1 |
fuses | x2 (including a spare contained in the fuse holder) |
interlock disable key(1505) | x1 |
tc receptacle adaptor box (1932) | x1 |
hipot test lead for ph / pl (1101) | x2 |
rs232 link cable 1.5m(1130) | x1 |
dut power cable 40a(1148) | x1 |
run module (opot.752) |
w | range : 0-1kw / 1-10kw resolution : 0.5w / 5w |
a | range : 0.000 !v 3.500 a / 3.00 - 40.00a , resolution : 0.001a / 0.01a |
pf | 0.000 -1.000 |
leakage current | range : 0.00 - 10.00ma , resolution : 0.01ma |
ac current high/ low limit setting | range: | 0-40a |
resolution: | 0.01a/step |
ac current display | range: | 0.000 !v 3.500 a / 3.00 - 40.00a |
resolution: | 0.001a / 0.01a |
accuracy : | ± (2 % of reading + 5 counts) / ± (2 % of reading + 2 counts) |
ac power high/ low limit setting | range: | 0 - 10kw |
resolution: | 1w |
ac power display | range: | 0.0 -1000.0w / 1000 - 10000w |
resolution: | 0.5w / 5w |
accuracy : | ±(2 % of reading + 2 counts ) |
power factor high/ low limit setting | range: | 0.000 - 1.000 |
resolution: | 0.001 |
power factor display | range: | 0.000 - 1.000 |
resolution: | 0.001 |
accuracy : | ± (2 % of reading + 2 counts ) |
leakage current high/ low limit setting | range: | 0.00 - 10.00ma |
resolution: | 0.01ma |
leakage current display | range: | 0.00 - 10.00ma |
resolution: | 0.01ma |
accuracy : | ± (2 % of reading + 2 counts ) |
delay timer setting | range: | 0, 0.2 - 999.9 |
resolution: | 0.1 |
dwell timer setting | range: | 0, 0.1-999.9 (0=continuous) |
resolution: | 0.1 |
timer display | range: | 0.0 - 999.9 seconds |
resolution: | 0.1 second |
accuracy: | ± (0.1% of reading + 0.05 seconds) |
power control | it is applied to control transformer box (1931/1931s) or ac power source (6700/6800/6400 series) |
hv & gb link module (opt. 753) |
max. acw 4000vac, dcw 4000vdc & max. ground bond 40a |
gpib interface card (opt. 731) |
printer port interface card (opt. 732) |
multi-function interface card (opt. 751) |
including usb port a type, combination of rs-232, rs-485 port, and bar code input with internal memory capability. |
high measurement range 35marms / 70mapeak & 4mds (opt. 754) |
35marms / 70mapeak for iec60990 fig4(u1 and u2), iec60990 fig5(u1 and u3), ul484, iec60601 test requiremt , notice : option this md circuit have to replace md module all md circuits . |
high measurement range 35marms / 70mapeak & 4mds (opt. 754) |
1931 series transformer box (500va/ 1kva/1.8kva/3.6kva) | multi-tape of 1.0, 1.06, 1.1 times of input voltage. |
1931-s series transformer box (500va/ 1kva/1.8kva/3.6kva) | multi-tape of 0.8, 0.85, 0.9, 1.0, 1.06, 1.1 times of input voltage. |
ac power source (option) add this option must optional opt.752 run module. |
6800 series digital ac power source (6805, 6810 , 6820 , 6830 and 6840 ) |
6700 series programmable ac power source (6705, 6710 , 6720 , 6730 and 6740 ) |
6400 series ac power source (6402, 6403, 6405, 6410) |